Home

 

UVS-E-T

UVS-E-T • Spectral sensitivities precisely adapted to the human skin

 • Very good cosine response: ±2.5% between 0º - 70º zenith   angle
                         
 • Excellent long-term stability


 

 

 



The UVS-E-T has a spectral response that is adapted to the erythemal (sunburn) action spectrum of the human skin (CIE 1987 weighting). Hence, this radiometer collects information of medical and public concern: UV Index, UV erythemal dose and dose rate.

The sensor output in MED/h (Minimum Erythemal Doses per hour) can be used to predict how long the human skin can be exposed to the sun before an erythema (reddening or sunburn) occurs. Due to the Robertson-Berger type spectral response of the UVS-E-T radiometer, this one can be related to existing monitoring network data. The high accuracy UVS-E-T is thermostated (heated / cooled at 25 ºC) including filters, detector and pre-amplifier. Due to the heating/cooling element it requires external power.

The UVS-E-T has a sealed, weather resistant case made from protected aluminum. The sensor permits continuous outdoor operation. The UVS-E-T can easily be adjusted in the horizontal by help of three leveling screws and the built-in circular bubble level.

For more information or to make an enquiry please contact us.

Product Specification

Spectral range 270 to 390 nm
Output range 0 to 0.6 W/m²
Response time 0.5 s
Non-linearity < 0.05 %
Accuracy > 95% (for daily sums)
Operating temperature range -25 to +50 °C
Supply voltage DC 7 to 18 V


Featured Products

CMP 3

VENTUS-UMB

WS500-UMB

Home Contact Us Site Map Terms & Conditions  

 